PSANA 6100
Gas analyzer


PSANA 6100 is a series of analyzer with high quality and high performance. Its sensor combined the leading technology, the measurement from it would be accurate and reliable.  

FEATURES:
• Easy operation • Maintenance free • high accuracy • Stable performance


 GAS Principle Range Code
 High purity O2 paramagnetic 98-100% PM1
 Constant O2 0-100% range optional PM2
 Trace O2 ECT 0-10,100,1000 ppm EC1
 Trace O2 ZRO ppm to 21% ZR1
 Ar/O2 TCD 0-100% range optional TCO
 Ar/N2 0-100% range optional TCN
 H2 0-100% range optional TCH
 CO2 IR ppm to % level IR
 CO ppm to % level IRCO
 CH4 ppm to % level IRCH4
 Dew point AL2O3 -100~20℃ AL2O3

SPECIFICATIONS:

 GAS High purity O2 Constant O2 Trace O2 Trace O2(ZRO)
 Range

98-100% O2
with pressure
compensation

0-100% O2
Range optional

0-10,100,1000
PPM O2

0-25%
0-10ppm O2 (minimum)
 Accuracy ±0.03% ±0.1% ±2% F.S <0.2ppm O2
 Response Time <10 S (T90) <10 S (T90) 21% TO 10 PPM<1 H <30 S (T90)
 Zero Drift ±0.05% /WEEK ±0.1% / WEEK ±2% F.S / WEEK ±2% F.S / WEEK
 Span Drift ±0.05% /WEEK ±0.1% / WEEK ±2% F.S / WEEK ±2% F.S / WEEK
 Inlet Pressure 5 psig ±3 psig 5 psig ±3 psig 5 psig ~30 psig 5 psig ±3 psig
 Flow 50 - 150 ml /min 50 - 150 ml /min 500–2500 ml/min 300 - 500 ml/min

 Gas Ar H2 CO2 CO CH4
 Range 0-100%
0-1%(minimum)
0-100%
0-1%(minimum)
0-100%
0-2000 ppm
0-50%
0-1%(minimum)
0-50%
0-1%(minimum)
 Accuracy <  ±1% <  ±1%
 Response Time < 5 S (T90) <6 S (T90)
 Zero Drift <  1% /WEEK < 2% FSD/WEEK
 Span Drift <  1% /WEEK < 2% FSD/WEEK
 Inlet Pressure 10 psig 5 psig±3 psig
 Flow 1000- 2000 ml/min. 50 - 100 ml /min.

 Output Signal 4-20mA isolated
 Display 0-100%0-1%(minimum)
 Connection IN:1/8'',OUT:1/4'',SWAGELOK
 Power 90VAC – 264 VAC / 47 – 63 Hz ; power< 50VA
 Work Temperature 5-45℃

Dimension:19’’,3U,360mm (W:440mm,H:133mm,D:360mm)